Module 869 has the following faults:
Channels 24-36 have a 2 sensor open (noise on neighbor due to high current)
Channel 66 Pinhole
Channel 87 Pinhole
Channels 99,100 2 sensor open
Channel 213 1 sensor open
Channel 216 Pinhole
Channels 233,235 Shorted
Channels 264,265 Shorted
Channel 324 Pinhole
Channels 340,341 Shorted
Channel 358 Pinhole
Channel 410 2 sensor open
Channel 471 2 sensor open
Channel 476 1 sensor open
Channel 477 2 sensor open
Channel 491,493 Shorted
All faults found and identified correctly other than the 340-341 short.
Channel 341 is marked as failing multiple tests, but 340 has fairly regular
responses except noise is slightly high.
Summaries of test results for UCSB and FNAL
IV curve
Module 881 has the following faults: Channel 40 Pinhole Channels 90-100 2 sensor opens Channel 146 Pinhole Channel 173 1 sensor open Channels 187,189 Shorted Channels 274-276 2 sensor open Channel 286 Pinhole Channels 292,293 Shorted Channel 318-319 1 sensor open Channel 406 1 sensor open Channel 434-444 2 sensor opens Channel 482 Pinhole All faults found and identified correctly. After testing at UCSB channels 434-436 and 442-444 pulled to improve performance around micro-discharge area, but was not re-tested prior to shipping to FNAL. Summaries of test results for UCSB and FNAL